  1. Technical Parameters
Item No. Parameter Name Technical parameters
1.1 Equipment Name Thermal Shock Test Chamber (Two-Chamber)
1.2 Equipment Introduction This series of products is used to test the tolerance level of material structures or composite materials under a continuous environment of extreme high and low temperatures in an instant. Thus, it can test the chemical changes or physical damage caused by thermal expansion and contraction within the shortest possible time. It is suitable for testing physical changes in fields such as electronics, automotive, metals, chemicals, building materials, communication components, national defense industry, aerospace industry, electronic chips (ICs), IT, and other related sectors.
1.3 Equipment Model BS-80HA
1.4 Effective Volume 80L
1.5 Internal Chamber Dimensions 500*400*400 (W*H*D)(mm)
1.6 External Dimensions 1200*2063*1550(W*H*D) (mm)
1.7 Temperature Range Heat Storage in the High-Temperature Zone:+60℃ ~ +170℃

Cold Storage in the Low-Temperature Zone: -10℃ ~ -55℃

1.8 Thermal Shock Range High-Temperature:60℃~150℃

Low-Temperature: -10℃~ -45℃

1.9 Thermal Shock and Recovery Time Thermal Shock Time: 10S

Temperature Recovery Time: 5 min

 

Shock Process (as shown in the figure)

Green represents the heat storage curve of the high-temperature zone

Red represents the shock curve of the test zone

Blue represents the cold storage curve of the low-temperature zone

1.10 Temperature Fluctuation ≤±0.5℃
1.11 Temperature Error ≤±2℃
1.12 Temperature Uniformity ≤±2℃(Temperature Uniformity is defined as the average of the difference between the measured maximum temperature and the measured minimum temperature in each test.)
1.13 Preheating Zone Temperature Rise Rate ≥3℃/min (Full-process Non-linear)
1.14 Precooling Zone Temperature Drop Rate ≥1℃ (Full-process Non-linear)
1.15 Functional Requirements (Air-based) In chamber Space Cooling and Heating

According to the temperature setting, the indoor space reaches the set temperature. It is divided into three parts: high-temperature storage zone, low-temperature storage zone, and test zone. The three zones are independent of each other, and temperature switching is achieved through air dampers.

1.16 Weight 450KG
1.17 Power 380V/22KW
1.18 Others Note & Prohibition for the Test Equipment

1. Note

The above performance indicators are values measured under the conditions of: ambient temperature of +25℃, relative humidity ≤ 85% RH, and no test samples loaded.

2. Prohibition

This test equipment is prohibited from the following:

· The testing or storage of samples of flammable, explosive, or volatile substances.

· The testing or storage of samples of corrosive substances.

· The testing or storage of biological samples.

· The testing or storage of samples with strong electromagnetic emission sources.

· The testing or storage of samples of radioactive substances;

· The testing or storage of samples that may generate flammable, explosive, volatile, highly toxic, corrosive, or radioactive substances during testing or storage.

 

  1. Compliant Test Standards
Item No. Standards Explanation of Standards
2.1 General standards GJB367.2-87 Thermal Shock Test
GJB360.7-87 Thermal Shock Test
GJB150.5-86 Thermal Shock Test
GB2423.22-87 Temperature Change Test Method
GB/T2423.2-2008(IEC60068-2-2:2007) High-Temperature Test Method Bb
GB/T2423.1-2008(IEC60068-2-1:2007) Low-Temperature Test Method Ab
2.2 Acceptance Execution Standards Manufacturing Standards:GB/T10586-2006,GB/T10592-2008
Verification Standards :GB/T5170.2-2017,GB/T5170.5-2016
Technical Standards: Acceptance shall be conducted in accordance with the signed technical index requirements
  1. Equipment Structure
Item No. Description Technical Specification Parameters
3.1 Cabinet Structure  

Two-Chamber Type: Preheating Zone, Precooling Zone, Mobile Basket (User Operation Area).

3.2 Shock Schematic Diagram
3.3 Cabinet Material Specifications Outer Wall Material: 1.2mm cold-rolled steel plate with electrostatic spraying

Inner Wall Material: High-quality heat-resistant and cold-resistant 1.0mm&304 stainless steel plate

Cabinet Insulation Material: 100mm glass fiber

3.4 Insulation Layer e zone and low-temperature zone is achieved using a thickened insulation layer, and thermal insulation for the connecting plate of the mobile basket’s moving hole is achieved using an epoxy resin plate.
3.5 Mobile basket The mobile basket consists of several components: a bottom heat-insulating end cover plate, a middle sample placement frame, a top heat-insulating end cover plate, counterweight blocks, and a basket lifting chain.

The overall frame is made of SUS304 stainless steel. High-temperature-resistant glass fiber cotton is embedded in the upper and lower end covers for heat insulation, which effectively prevents cross-temperature transfer and avoids affecting the refrigeration effect.

The sample rack frame adopts a perforated plate structure, and a sample rack height-adjustable indexing frame is longitudinally equipped. It can adjust the spacing of the sample racks without blocking airflow circulation, ensuring temperature uniformity inside the chamber.

Counterweight blocks are embedded in the bottom end cover of the basket to ensure good sealing during low-temperature testing.

3.6 Cabinet Door The high-temperature zone and low-temperature zone are each equipped with a separate door, both adopting a single-door design. Explosion-proof handles are used, and silicone sealing strips are installed to ensure the cabinet doors are sealed without leakage.
3.7 Test Sample Rack 2 Pieces of SUS304 Stainless Steel Grid-Type Storage Racks (Load Capacity ≤ 40kg per Piece)
3.8 Air Circulation Duct A multi-blade centrifugal circulation fan is adopted, with a reinforced shaft center and high/low temperature-resistant rotating blades made of aluminum alloy, to achieve forced convection.

Air supply mode: Horizontal diffusion + vertical heat exchange + arc-shaped circulation.

The high/low temperature air supply damper switching is driven by a high-speed air cylinder under automatic program control.

3.9 Power Distribution and Control Cabinet Main Power Circuit Breaker,Controller,Power Distribution Board,Cooling Fan,Over-Temperature Protector,Air Cylinder,Air Pressure Protector etc.
3.10 Refrigeration Mechanical Chamber Internally Installed Refrigeration Units and Accessories
3.11 Others LCD Touch-Type Programmable Controller, Basket Initial Position Changeover Switch, Main Power Switch, Running Indicator Light, Abnormal Indicator Light,Completion Indicator Light, Defrosting Indicator Light, Power Indicator Light, USB Communication Interface
  1. Circulation System
Item No. Description Technical Specification Parameters
4.1 Air Outlet Mode It adopts a forced internal circulation mode, with left-side air outlet and right-side air return switched via air dampers. The air outlet guide vanes can disperse cold and hot air throughout the entire chamber interior, ensuring uniformity and stability.

4.2 Circulation Motor It is a long-shaft motor, installed outside the chamber, avoiding the influence of temperature and humidity inside the chamber. The long shaft extends directly into the interior of the chamber, with silicone insulation and sealing.
4.3 Centrifugal Impeller It is a multi-blade centrifugal circulating aluminum wind wheel, featuring light weight, large air volume, and high air pressure. Additionally, it exhibits resistance to high and low temperatures, a small expansion coefficient, and low operating noise.
4.4 Circulating Air Duct The circulating air duct is matched with the refrigeration and heating system, and the sealing plate is closely attached to the evaporator and heating tube, allowing the circulating air to fully absorb the cold energy from the evaporator and the heat energy from the heating tube.
4.5 Sensor Class A Platinum Resistance PT100+T Temperature and Humidity Sensor, featuring stable operation and accurate detection.
  1. Refrigeration System
Item No. Description  Technical Specification Parameters
5.1 Refrigeration Mode Compressor-Based Refrigeration
5.2 Compressor French “Tecumseh” Hermetic Compressor
5.3 Evaporator High-Efficiency Multi-Stage Fin-and-Tube Evaporator with Hydrophilic Coating (Thickened Fin Type)
5.4 Condenser Shell-and-Tube Condenser (Air-Cooled Type)
5.5 Refrigerant Fluoride-Free Eco-Friendly Refrigerant, DuPont (USA) R404A
5.6 Thermal Expansion Valve The core control logic of the thermal expansion valve is: The opening degree of the expansion valve is controlled by the superheat degree of the gaseous refrigerant at the evaporator outlet, thereby adjusting the flow rate of refrigerant entering the evaporator.
5.7 Oil Separator The core function of the oil separator is to separate the refrigerating oil and refrigerant, allowing each substance to perform its respective function as required and preventing oil films from reducing heat exchange efficiency.
5.8 Filter-Drier The filter-drier consists of two parts: the filter and the dryer. The filter is used to block solid particles such as iron filings, welding slag, and dirt, preventing them from clogging the passages; the dryer is used to absorb moisture circulating with the refrigerant, avoiding “ice blockage” inside the pipelines.
5.9 Solenoid Valve Danfoss Solenoid Valve (for coarse adjustment of the cooling capacity of the entire refrigeration system)
5.10 Leak Detection Method Nitrogen purging and pressure holding for 24 hours is adopted for leak detection testing of all pipelines in the entire system.
5.11 Vibration Damping Design Compressor Base and Vibration Isolation Treatment: The base features a water-leakage prevention design. It is a self-evaporating waterproof base that effectively prevents condensed water from leaking out. Vibration isolation treatment is applied to moving components such as the compressor and fan, achieving a low-noise design.
  1. Heating System
Item No. Description Technical Specification Parameters
6.1 Heating Method It adopts an imported SSR (Solid-State Relay) for driving, featuring contactless control, no noise, no electric arc, and stable reliability.
6.2 Heating Tube It adopts a stainless steel finned nickel-chromium wire high-speed heater.
6.3 Heating System and Electrical Components It is a completely independent system that does not affect the refrigeration and humidification systems. Temperature control is achieved through microcomputer calculation and PID automatic adjustment; the output power changes based on the comparison result between the actual temperature value and the set temperature value, so as to achieve high precision, high efficiency, and high stability. Major low-voltage electrical components all adopt products from well-known brands: “Schneider” AC contactors and thermal overload relays, “OMRON” small intermediate relays, and Delixi signal lamps, switches, and circuit breakers.
  1. Control System
Item No. Description Technical Specification Parameters
7.1 Control Method Color Touch Screen + PLC Control System
7.2 Screen Display 7-inch screen with a resolution of 800*480, 65535 true colors, and LED backlight display; direct display of temperature and humidity set values (SV) and actual values (PV); Chinese/English language switchable display with true-color touch input; capable of displaying execution program number, segment number, remaining time, cycle times, and operation time; program editing and graphic curve display.
7.3 Temperature and Humidity setting Set it independently according to the test temperature range requirements.

7.4 Data Logging Built-in recording program, the memory in the controller can store 3 months of 24hour operation data. It can automatically generate temperature and humidity curve files, data, operation records, alarm records, etc., and can display set values and actual values, cycle numbers, temperature segments, temperature values, time, etc. in both digital and graphical forms.
7.5 External Connection Communication methods such as USB, RS485, and LAN ports are supported, enabling external connection to a computer for full-function control, data storage, printing, and recording.
7.6 Running Mode Rated value/program/slope running mode.
7.7 Input Method RT100×2 (dry bulb and wet bulb) / 4~20mA, both built-in simultaneously
7.8 Output Method Temperature×1, Humidity×1, END×1, RUN×1, T1~T4, TROUBLE×1, H1×1, TIME SIGNAL×3.
7.9 Display Precision Temperature:0.01℃

Humidity:0.01%RHT

Time:0.01 seconds

7.10 System Capacity 1~120 test programs (i.e., can be divided into 120 program groups), maximum 1~1200 steps (i.e., 1200 action steps), and single-step time up to 9999 hours 59 minutes.
7.11 Login Method Chinese/English Name Login, Chinese/English Test Login, Operation Time Display.
7.12 Deviation Correction It prevents parameter deviations caused by various factors; after correction, the temperature and humidity values can be more accurate and standardized.
7.13 System Functions Sequence Control, Reservation Control, Power-off Recovery, Fault Display.
7.14 Password Control Multi-level password protection can be set to prevent unauthorized use by non-operators and unauthorized modification of internal parameters.