BS-B-8000AD Walk-In Constant Temperature and Humidity Test Chamber-1005

  1. Technical Parameters
Item No. Parameter Name Technical parameters
1.1 Equipment Name Walk-In Constant Temperature and Humidity Test Chamber
1.2 Equipment Introduction The floor-standing constant temperature and humidity laboratory is a macroclimate simulation device designed for testing the aging performance of electrical and electronic products. It is suitable for evaluating thermal and humidity resistance in industries including new energy, automotive

manufacturing, home appliances, semiconductors, electronics, aerospace, academic institutions, and instrumentation components. The system also conducts adaptability tests under extreme environmental conditions such as high temperatures, low temperatures, humid heat, and alternating humid-heat cycles.

Product features 1、Using multi-channel, environmentally friendly and energy-saving (low noise) mechanical compression refrigeration system.

All materials, components, and raw materials for this equipment, whether domestically produced or imported, are eco-friendly products. Based on the compression refrigeration characteristics of machinery, a dual-stage multi-way main-bypass system is adopted. The multiple main paths and bypasses can be automatically selected according to different operating conditions. The entire

control system operates autonomously through intelligent control based on actual working conditions, replacing the traditional method of using thermostats to regulate heaters to offset cooling capacity. This achieves energy-saving effects by reducing energy consumption to just 80% of similar products. It also decreases the operational load on refrigeration compressors,minimizes equipment vibration and noise, and improves control precision. The refrigeration compressor’s output adapts to various temperature change rates required by laboratories while controlling power consumption to enhance energy efficiency. Noise reduction measures are implemented throughout the structure and mechanical processes using sound absorption and noise elimination techniques, effectively minimizing overall equipment noise.

2. Self-developed multi-way and multi-stage throttling system (refrigeration system).

The refrigeration system of this equipment is designed with different main and bypass capillary roads according to different working conditions of the system. The intelligent control system automatically controls the capillary throttle system according to working conditions, and automatically adjusts the flow size to achieve the purpose of temperature change, and the temperature field changes uniformly with minimal temperature fluctuation.

3. The broadband pressure balanced temperature control system is applied to realize rapid temperature control through the fast-converging PID adaptive algorithm.

By integrating a temperature conditioning chamber, static pressure chamber,and centrifugal fan-driven forced air duct system, we ensure efficient heat exchange while delivering uniform airflow through the overhead perforated plate to every workspace (experimental area), achieving optimal temperature distribution. To accelerate thermal stabilization, our company collaborated with Weilun Corporation to develop a proprietary PID adaptive control algorithm. This intelligent control method continuously adjusts parameters

based on real-time temperature data, ensuring rapid response, minimal overshoot, and highly stable performance throughout the process.

4、Realize remote visual control.

The equipment features a human-machine interface and a touch-mode automated control system for seamless operation. It connects to remote computers via an RS485 communication interface, enabling network management and real-time monitoring. The system’s graphical interface

simplifies control operations, provides automatic fault alerts, displays energy-saving suggestions with visual indicators, and guides users through quick troubleshooting solutions.

1.3 Equipment Model BS-B-8000AD
1.4 Effective Volume 8000L
1.5 Internal Chamber Dimensions 2000*2000*2000 (W*H*D)(mm)
1.6 External Dimensions 3250*2400*2350(W*H*D) (mm)

( Subject to the final confirmed 3D drawing )

1.7 Temperature Range -40℃~150℃
1.8 humidity range 20~98%R. H

(Temp. & Humidity Comparison Table)

1.9 Heating Rate -40℃ ~ 150℃ ≥3℃/min

(Full-process Non-linearity & No-load Acceptance)

1.10 Cooling Rate 150℃ ~ -40℃ ≥0.7~1℃/min

(Full-process Non-linearity & No-load Acceptance)

1.11 Temperature and Humidity Display Accuracy Temperature:0.01℃

Humidity:0.1%R.H

1.12 Temperature and Humidity Deviation Temperature:±2℃

Humidity: When ≥75%RH, the humidity error shall be ≤+2 to -3%RH

Humidity: When <75%RH, the humidity error shall be ≤+5%RH

1.13 Temperature and Humidity Fluctuation Temperature:≤±0.5℃

Humidity:+2/-3%R.H

1.14 Temperature and Humidity Uniformity Temperature:≤2℃

Humidity:+2/-3%R.H

1.15 Cooling Method Water Cooling
1.16 Weight 2300KG
1.17 Power 380V/32.5KW
1.18 Others All the above performance indicators refer to the data measured under the conditions of no load, no test sample, and ambient temperature of 25±2℃.

Operating Noise:70dB (measured at 1 meter from the chamber door and 1.5 meters in height).

  1. Compliant Test Standards
Item No. Standards Explanation of Standards
2.1 General standards GB/T2423.2-2008

Test Method: High Temperature

GB/T 2423.1-2008

Test method: Low Temperature

GJB150.4A-2009

Low Temperature Test Method

GJB150.3A-2009

High Temperature Test Method

G/BT 2423.4-2008/IEC6008-2-30:2005

Test Db: Alternating Damp Heat Method

GBT 2424.6-2006

Environmental Testing for Electrical and Electronic Products – Performance Qualification Report of Temperature and Humidity Test Chambers

GB/T10586-2006

Technical Specifications for Damp Heat Test Chambers

GBT 2424.5-2006

Environmental Testing for Electrical and Electronic products – Performance Qualification Report of  Temperature Test Chambers.

G/BT 2423.4-2008/IEC6008-2-30:2005 test Db: alternating wet heat method
2.2 Acceptance Execution Standards Manufacturing Standards:GB/T10586-2006,GB/T10592-2008
Verification Standards :GB/T5170.2-2017,GB/T5170.5-2016
Technical Standards: Acceptance shall be conducted in accordance with the signed technical index requirements
  1. Equipment Structure
Item No. Description Technical Specification Parameters
3.1 Cabinet Structure The split frame structure is welded with national standard channel steel, which ensures the overall reliability and safety performance of the equipment with strong durability, low vibration and low noise during operation.
3.2 Inner Chamber Material It adopts 1.0mm-thick SUS #304 stainless steel.

l; the whole full welding treatment is adopted to ensure the sealing performance of the inner box, and the bottom reinforcement design ensures the bearing capacity of the bottom ≥500KG/㎡ average load.

3.3 Outer Shell Material It adopts cold-rolled steel plates with a thickness of 1.2mm, featuring double-sided spraying treatment and a blue-white color(Customizable according to customer requirements, color samples or color codes are available upon request).
3.4 Thermal Insulation Material The thermal insulation layer has a thickness of 100mm and uses fire-resistant, high-strength rigid polyurethane foam insulation material. Thermal insulation coefficient: less than 0.0212 kcal/(m・hour).
3.5 Cabinet Door Structure The double door is sealed with double-layer silicone rubber between the door and the box, which can withstand high and low temperature and resist aging to ensure the sealing performance. The door frame and edge are equipped with electric heating device to prevent condensation and frost, so as to avoid condensation on the surface of the box.
3.6 Test Observation Window Set 1 vacuum multi-layer explosion-proof glass window, the window is designed to be frostproof: the window is coated with heating to prevent frost, so as to prevent water condensation; a safe high-brightness energy-saving lighting lamp is set above the observation window (which can be manually switched on and off).
3.7 Air Duct Components It adopts a circulating fan with a motor, an aluminum extended shaft, and a high/low temperature-resistant aluminum alloy multi-blade circulating fan, achieving forced convection vertical diffusion circulation. The air outlet is adjustable.
3.8 Test Port One test port with a diameter of 100mm is reserved on the left side of the chamber, and one set of silicone plug + sealing cover is provided.
3.9 Three-Color Indicator Light Green for normal operation, yellow for standby, and red for fault alarm.
3.10 Drainage Trough Drainage troughs are provided around the interior of the cabinet, facilitating the discharge of condensed water and preventing internal water accumulation.
3.11 Balance Hole A pressure balance hole is installed inside the chamber to balance the pressure between the inside and outside of the chamber.
3.12 Swivel Caster Swivel casters are installed at the bottom of the equipment, facilitating the relocation, movement and positioning of the equipment.
Assembly diagram

  1. Circulation System
Item No. Description Technical Specification Parameters
4.1 Air Supply Method It adopts a forced internal circulation mode with top air supply and bottom air return. The air deflector at the air outlet can disperse cold and hot air to the entire interior of the chamber, ensuring uniform and stable distribution.
4.2 Circulation Motor It is a long-shaft motor, installed outside the chamber, avoiding the influence of temperature and humidity inside the chamber. The long shaft extends directly into the interior of the chamber, with silicone insulation and sealing.
4.3 Centrifugal Impeller It is a multi-blade centrifugal circulating aluminum wind wheel, featuring light weight, large air volume, and high air pressure. Additionally, it exhibits resistance to high and low temperatures, a small expansion coefficient, and low operating noise.
4.4 Circulating Air Duct The circulating air duct is matched with the refrigeration and heating system, and the sealing plate is closely attached to the evaporator and heating tube, allowing the circulating air to fully absorb the cold energy from the evaporator and the heat energy from the heating tube.
4.5 Sensor Class A Platinum Resistance PT100+T Temperature and Humidity Sensor, featuring stable operation and accurate detection.
  1. Refrigeration System
Item No. Description  Technical Specification Parameters
5.1 Refrigeration Mode Compressor-Based Refrigeration
5.2 Compressor German “Bizer” fully enclosed compressor
5.3 Evaporator High-Efficiency Multi-Stage Fin-and-Tube Evaporator with Hydrophilic Coating (Thickened Fin Type)
5.4 Condenser Shell-and-Tube Condenser (Water-Cooled Type)
5.5 Refrigerant Fluoride-Free Eco-Friendly Refrigerant, DuPont (USA) R404A
5.6 Thermal Expansion Valve The core control logic of the thermal expansion valve is: The opening degree of the expansion valve is controlled by the superheat degree of the gaseous refrigerant at the evaporator outlet, thereby adjusting the flow rate of refrigerant entering the evaporator.
5.7 Oil Separator The core function of the oil separator is to separate the refrigerating oil and refrigerant, allowing each substance to perform its respective function as required and preventing oil films from reducing heat exchange efficiency.
5.8 Filter-Drier The filter-drier consists of two parts: the filter and the dryer. The filter is used to block solid particles such as iron filings, welding slag, and dirt, preventing them from clogging the passages; the dryer is used to absorb moisture circulating with the refrigerant, avoiding “ice blockage” inside the pipelines.
5.9 Solenoid Valve Danfoss Solenoid Valve (for coarse adjustment of the cooling capacity of the entire refrigeration system)
5.10 Leak Detection Method Nitrogen purging and pressure holding for 24 hours is adopted for leak detection testing of all pipelines in the entire system.
5.11 Vibration Damping Design Compressor Base and Vibration Isolation Treatment: The base features a water-leakage prevention design. It is a self-evaporating waterproof base that effectively prevents condensed water from leaking out. Vibration isolation treatment is applied to moving components such as the compressor and fan, achieving a low-noise design.
  1. Heating System
Item No. Description Technical Specification Parameters
6.1 Heating Method It adopts an imported SSR (Solid-State Relay) for driving, featuring contactless control, no noise, no electric arc, and stable reliability.
6.2 Heating Tube It adopts a stainless steel finned nickel-chromium wire high-speed heater.
6.3 Heating System and Electrical Components It is a completely independent system that does not affect the refrigeration and humidification systems. Temperature control is achieved through microcomputer calculation and PID automatic adjustment; the output power changes based on the comparison result between the actual temperature value and the set temperature value, so as to achieve high precision, high efficiency, and high stability. Major low-voltage electrical components all adopt products from well-known brands: “Schneider” AC contactors and thermal overload relays, “OMRON” small intermediate relays, and Delixi signal lamps, switches, and circuit breakers.
  1. Humidification System
Item No. Description Technical Specification Parameters
7.1 Humidification Method It is a fully automatic humidification module system with integrated closed-loop humidity control, designed and manufactured using advanced technology. It is an improved and upgraded multi-functional integrated humidification system,featuring superior characteristics such as fast humidification, high efficiency, ultra-low failure rate, and higher reliability. It adopts SSR-driven contactless control and intelligent PID automatic calculation to adjust output power.
7.2 Water Level Control It features automatic water level control and automatic water filling functions, which make the humidification system more efficient, the control more stable, and the equipment operation more reliable. These functions avoid the drawback of high failure rates in humidification systems and overall improve the equipment’s reliability.
7.3 Humidifier Closed-body structure, easy to install, can be installed on the outside of the room, the system structure is simpler, more efficient and practical.
7.3 Humidity sensor Electronic temperature and humidity sensor: Select imported electronic temperature and humidity sensor.